A complete reference across all six domains of haemodynamic monitoring — CVP to microcirculation, static indices to dynamic, invasive to non-invasive. Six modules. Updated 2025–26.
The oldest continuous haemodynamic monitoring tool in clinical use — measuring pressure in the right atrium to reflect right ventricular filling.
What is CVP?
CVP is the pressure measured in the superior vena cava (SVC) or right atrium (RA). It reflects right ventricular end-diastolic pressure (RVEDP) and serves as a surrogate for right ventricular preload. Measured in mmHg (or cmH₂O — divide by 1.36 to convert).
Normal CVP
2–8
mmHg
in cmH₂O
3–10
cmH₂O
Low — hypovolaemia
<2
mmHg
Elevated — overload / RHF
>10
mmHg
Sites of Insertion
Site
Advantages
Disadvantages
Infection Risk
Internal Jugular (R)
Straight path to RA, high success rate
Carotid artery risk, uncomfortable
Low
Subclavian
Most comfortable, lowest infection risk
Pneumothorax (1–3%), cannot compress
Lowest
Femoral
Easy in emergency, no PTX risk
High infection risk, less accurate CVP
Highest
External Jugular
Visible, accessible
Tortuous, may not reach SVC
Low
✦ Tip Placement
Optimal tip position: cavo-atrial junction — lower third of SVC, at or just above the carina on CXR.
The CVP Waveform
Five components — three positive waves (a, c, v) and two negative descents (x, y). Memorise these cold.
▸ CVP Waveform — Two Complete Cycles, Annotated
a · c · v
aAtrial contraction — after P wave — absent in AF
cClosure of tricuspid valve + RV isovolumetric contraction — after QRS
vVenous filling of atrium while TV is shut — after T wave
xeXit — atrial relaxation + TV annular descent (systolic dip)
yemptying Yields to RV — TV opens, RV starts to fill (diastolic dip)
Pathological Waveform Changes
Pattern
Mechanism
Condition
Giant a wave
Atrium contracts against closed/resistant TV
Pulmonary hypertension, tricuspid stenosis, RV failure, AV block
Cannon a wave
Atrium contracts against closed TV (AV dissociation)
Complete heart block, VT, junctional rhythm
Giant v wave
Regurgitant flow into atrium during systole
Severe tricuspid regurgitation
Absent a wave
No organised atrial contraction
Atrial fibrillation
Steep x, blunted y — "M" pattern
Pericardial fluid limits ventricular filling; x preserved, y obliterated
Cardiac tamponade
Steep x AND steep y — "W" pattern
Both descents prominent; rapid early filling then abrupt halt
Constrictive pericarditis
Absent / blunted x
No atrial relaxation visible
Tricuspid regurgitation
Zero Reference & Transducer Positioning
Zero reference: Phlebostatic axis — intersection of 4th ICS and mid-axillary line (approximates RA level)
Patient can be supine or up to 30–45° HOB (if transducer is re-levelled)
Always measure at end-expiration — intrathoracic pressure closest to atmospheric
On mechanical ventilation: read at end-expiration = highest point on the CVP waveform during the respiratory cycle
Each 2.54 cm (1 inch) transducer error = ~2 mmHg reading error
⚠ The Central Limitation
CVP is a poor predictor of fluid responsiveness. Marik et al. 2008 meta-analysis (24 studies, 803 patients): CVP AUC-ROC = 0.55 — no better than chance. A single CVP value should never be used in isolation to guide fluid therapy.
✦ CLABSI Bundle
Maximal sterile barrier · Chlorhexidine skin prep · Subclavian or IJV preferred over femoral · Daily review of necessity · Hand hygiene. Subclavian has lowest infection rate.
Module 02 of 06
Pulmonary Artery Catheterisation
The Swan-Ganz catheter — balloon-tipped, flow-directed, and capable of measuring pressures through the entire right heart and into the pulmonary vasculature.
Historical note
Introduced by Swan and Ganz in 1970. A 7.5 Fr catheter that floats through the right heart when its balloon is inflated, guided by blood flow.
RA pressure
2–8
mmHg
RV systolic / diastolic
15–30 / 0–8
mmHg
PA systolic / diastolic
15–30 / 6–12
mmHg
mPAP
9–18
mmHg
PCWP (wedge)
6–12
mmHg
Cardiac Output
4–8
L/min
CI
2.5–4.0
L/min/m²
Mixed SvO₂
60–80
%
SVR
800–1200
dynes·s/cm⁵
PVR
20–120
dynes·s/cm⁵
PAC Ports & Lumens
Port
Location
Purpose
Proximal (blue)
RA — 30 cm from tip
CVP, CO injectate, drug/fluid infusion
Distal (yellow)
PA tip
PAP measurement, mixed SvO₂ sampling, PCWP when wedged
Balloon inflation
Tip
Inflate 1.5 mL air to float/wedge — never use liquids
Thermistor connector
4 cm from tip
Temperature sensing for thermodilution CO
VIP port (white)
Mid-catheter (RV position)
Additional infusion lumen
Waveform Recognition During Insertion
▸ RA → RV → PA → PCWP — pressure sequence on insertion
⚑ RV vs PA — the critical distinction
RV: high systolic + diastolic near zero, NO dicrotic notch · PA: high systolic + diastolic elevated (6–12) + dicrotic notch present. If you see high systolic + low diastolic with no notch — you're still in the RV. Advance further before deflating the balloon.
PCWP — Pulmonary Capillary Wedge Pressure
Balloon inflated in a branch PA occludes flow, creating a static column from catheter tip → pulmonary capillaries → pulmonary veins → left atrium. Therefore PCWP ≈ LAP ≈ LVEDP (in the absence of mitral stenosis or pulmonary venous hypertension).
Transpulmonary Gradient
TPG = mPAP − PCWP
Normal <12 mmHg · Elevated in pre-capillary pulmonary hypertension
✦ Clinical Pearl
High PEEP can convert Zone III to Zone II → suspect if PCWP fluctuates with ventilation or PCWP > PA diastolic. Lateral CXR can confirm tip position below LA.
Thermodilution Cardiac Output — Stewart-Hamilton
Inject a known volume of cold or room-temperature saline into the proximal (RA) port. The thermistor at the PA tip measures temperature change over time. Area under the temperature-time curve is inversely proportional to cardiac output.
SvO₂ is sampled from the distal (PA) port. VO₂ is often assumed (~125 mL/min/m²) rather than measured — a limitation. Requires a metabolic cart for true VO₂ measurement.
Derived Haemodynamic Parameters
Parameter
Formula
Normal
Units
SVR
(MAP − CVP) / CO × 80
800–1200
dynes·s/cm⁵
PVR
(mPAP − PCWP) / CO × 80
20–120
dynes·s/cm⁵
Stroke Volume
CO / HR × 1000
60–100
mL/beat
SVI
CI / HR × 1000
33–47
mL/beat/m²
DO₂ (O₂ delivery)
CO × CaO₂ × 10
520–720
mL/min/m²
VO₂ (O₂ consumption)
CO × (CaO₂−CvO₂) × 10
100–180
mL/min/m²
O₂ Extraction Ratio
VO₂ / DO₂ × 100
22–30
%
Haemodynamic Profiles in Shock
Shock Type
CO
SVR
PCWP
CVP
SvO₂
Cardiogenic
↓↓
↑↑
↑
↑
↓
Hypovolaemic
↓
↑
↓
↓
↓
Distributive (Septic)
↑/N
↓↓
↓/N
↓/N
↑ early
Obstructive (PE/Tamponade)
↓
↑
↓/N
↑↑
↓
Neurogenic
↓/N
↓↓
N
N/↓
↑/N
Complications of PAC
Arrhythmias: PVCs, VT/VF as catheter passes through RVOT — most common during insertion
RBBB (3–6%): catheter irritates right bundle branch; critical if pre-existing LBBB → complete heart block. Have pacing ready
PA rupture (0.02–0.2%): rare but life-threatening. Higher risk with PHT, elderly, anticoagulation. Caused by balloon over-inflation or inflation in a small-calibre vessel
PA Rupture — Emergency Management
Trendelenburg position · 100% O₂ · Lung isolation (double lumen tube, affected side dependent) · Bronchoscopy · Catheter balloon as temporary tamponade · Interventional radiology (embolisation) or surgical repair.
⚠ The PAC Controversy
PACMAN, ESCAPE, and FACTT trials found PAC-guided therapy did not improve mortality vs standard management in ICU patients, septic shock, or acute heart failure. PAC use has declined significantly since 2000. Modern practice: target-specific goals with the least invasive method that answers the clinical question.
Module 03 of 06
Cardiac Output Monitoring — Invasive Methods
Pulse contour analysis and transpulmonary thermodilution — comprehensive haemodynamic data without a PA catheter.
Pulse Contour Analysis — Core Principle
Derives CO by analysing the morphology of the arterial pressure waveform. The area under the systolic portion of the waveform is proportional to stroke volume, adjusted for estimated aortic compliance and impedance.
Core Relationship
SV ∝ Area under systolic portion of arterial waveform (corrected for aortic impedance / compliance)
System
Calibration
Access Required
Unique Features
PiCCO (Pulsion)
Transpulmonary thermodilution — required
CVC + radial or femoral arterial line
ITBV, EVLW, GEDV, CFI, GEF, PVPI
LiDCO plus
Lithium dilution — required
Peripheral IV + arterial line
Contraindicated on lithium therapy, early pregnancy
LiDCOrapid
Uncalibrated (nomogram)
Peripheral arterial line
Continuous SV/CO; good for trending and fluid responsiveness
FloTrac / Vigileo (Edwards)
Uncalibrated (auto-calibrated)
Any arterial line
Uses SD of pulse pressure + patient demographics; no external calibration needed
MOSTCARE / Argos
Uncalibrated
Arterial line
dP/dt-based analysis
PiCCO — Volumetric Parameters
PiCCO calibrates via transpulmonary thermodilution (TPTD): cold injectate via CVC proximal port; thermistor in femoral or radial artery line. Provides unique volumetric preload indices unavailable from PAC.
ITBVI — Intrathoracic Blood Volume
850–1000
mL/m²
GEDVI — Global End-Diastolic Volume
680–800
mL/m²
EVLWI — Extravascular Lung Water
<10
mL/kg
PVPI — Pulmonary Vasc Permeability
1.0–3.0
ratio
CFI — Cardiac Function Index
4.5–6.5
/min
GEF — Global Ejection Fraction
25–35
%
✦ EVLW in Clinical Practice
EVLWI >10 mL/kg = pulmonary oedema · >15 mL/kg = severe · PVPI >3 = increased permeability → ARDS pattern (vs hydrostatic oedema). This distinction directly guides de-resuscitation decisions and diuretic therapy.
Calibrated systems (PiCCO, LiDCO): more accurate but require recalibration every 8h or after major haemodynamic shifts
Uncalibrated systems: adequate for trending and relative changes, not absolute values
Module 04 of 06
Cardiac Output — Minimally & Non-Invasive Methods
Measuring CO without arterial cannulation or central access — the direction of modern haemodynamic monitoring.
Oesophageal Doppler Monitoring (ODM)
A probe passed into the oesophagus (in intubated, sedated patients) measures blood flow velocity in the descending thoracic aorta using continuous-wave Doppler. Aortic cross-sectional area is estimated from a nomogram to derive CO.
Devices: CardioQ (Deltex), HemoSonic
FTc (flow time corrected): normal 330–360 ms; ↓ in hypovolaemia; ↑ in vasodilation
Peak velocity (PV): reflects contractility; ↓ in poor ventricular function
Measures ~70% of CO (descending aorta); assumes a fixed proportion to upper body
Validated in perioperative goal-directed therapy (GDT) — reduced complications in major surgery
Limitations: requires sedation, frequent probe repositioning, oesophageal pathology a contraindication
Echocardiography — LVOT Doppler
SV via LVOT (Doppler)
SV = VTILVOT × π × (DLVOT/2)² → CO = SV × HR
VTI = velocity time integral (PW Doppler in LVOT, apical 5-chamber) · D measured in parasternal long-axis view
Most validated non-invasive CO method; within ±10–15% of thermodilution
Key source of error: LVOT diameter (squaring effect — small errors amplify significantly)
TEE: continuous intraoperative use (semi-invasive); TTE limited by ventilation and body habitus
PI = perfusion index from SpO₂ probe · PVI >14% in ventilated patients predicts fluid responsiveness (similar performance to PPV)
Module 05 of 06
Markers of Fluid Responsiveness
Static and dynamic indices — predicting which patients will increase cardiac output in response to a fluid challenge. The central question in resuscitation.
⚑ The Core Question
Only ~50% of haemodynamically unstable ICU patients are truly fluid-responsive. Fluid in a non-responder causes harm — pulmonary oedema, dilutional coagulopathy, abdominal compartment syndrome. We must predict before we pour.
Definition: Fluid responsiveness = ≥10–15% increase in stroke volume or CO after a fluid challenge.
Static Indices — Poor Predictors
Static filling pressures tell you where you are on the x-axis. They do not tell you the slope of the Frank-Starling curve — which is the only thing that matters for predicting response.
Index
Normal
AUC-ROC
Why It Fails
CVP
2–8 mmHg
~0.55 (no better than chance)
Does not reflect ventricular compliance or contractility
PCWP
6–12 mmHg
~0.56
Same fundamental limitation; more invasive
ITBVI (PiCCO)
850–1000 mL/m²
~0.64
Volumetric; better than pressure but still static
GEDVI (PiCCO)
680–800 mL/m²
~0.66
Reasonable preload estimate; not a prediction of response
Why Dynamic Indices Work — The Mechanism
Positive pressure inspiration → ↑ intrathoracic pressure → ↓ venous return → ↓ RV preload → (after 2-beat pulmonary transit time) → ↓ LV filling → ↓ SV. In a heart sitting on the steep part of the Starling curve (preload-dependent), this cyclical variation is exaggerated. In a heart on the flat part, it is minimal. This is the basis of all dynamic indices.
Occlude expiratory circuit for 15 seconds → prevents cyclic reduction in venous return
CO increase ≥5% → fluid responsive (AUC ~0.93)
Requires continuous CO monitoring; works only in controlled MV patients
Tidal Volume Challenge
Temporarily increase Vt from 6 → 8 mL/kg IBW for one minute
If PPV increases by >3.5% → would be PPV-positive at standard Vt (AUC ~0.97)
Useful when lung-protective low-Vt ventilation makes PPV unreliable
Comparison Table — Dynamic Indices
Test
Threshold
AUC-ROC
Spont. Breathing
AF
PPV
>13%
~0.94
No
No
SVV
>13%
~0.84
No
No
IVC Distensibility
>18%
~0.89
No
Limited
IVC Collapsibility
>50%
~0.75
Yes
Limited
PLR + CO measurement
≥10% ↑CO
~0.95
Yes
Yes
EEOT
≥5% ↑CO
~0.93
No
Limited
Tidal Volume Challenge
ΔPPV >3.5%
~0.97
No
No
✦ Clinical Bottom Line
In a mechanically ventilated patient with no spontaneous effort, regular rhythm, Vt ≥8 mL/kg: use PPV. In everyone else: use PLR + continuous CO measurement. EEOT adds value when you need confirmation in MV patients.
Frank-Starling Curve — The Visual
Module 06 of 06
Markers of Microvascular Perfusion
Global haemodynamics can be normalised while the microcirculation remains in crisis — the invisible failure that drives organ dysfunction.
The Macro-Micro Disconnect
A patient may have MAP ≥65, CO normal, CVP normalised — and yet have stopped, intermittent, and heterogeneous capillary flow at the microvascular level. This persistent microcirculatory failure is an independent predictor of organ dysfunction and mortality in septic shock.
Lactate & Lactate Clearance
Lactate Level
Interpretation
Action
<2 mmol/L
Normal
Reassuring; continue monitoring
2–4 mmol/L
Moderate — possible hypoperfusion or impaired clearance
Investigate; treat cause; serial measurement
>4 mmol/L
Severe — high mortality risk · Sepsis-3 criterion for septic shock
✦ Lactate is Not Only Anaerobic
Aerobic sources in sepsis: Warburg effect (aerobic glycolysis in inflammatory cells) · Exogenous epinephrine → β₂-stimulation → ↑ skeletal muscle glycolysis · Impaired hepatic clearance · Thiamine deficiency · Metformin toxicity · Respiratory muscle fatigue (severe asthma). Always interpret in context.
ScvO₂ — Central Venous Oxygen Saturation
Measured from the distal port of a CVC in the SVC. Reflects the balance between O₂ delivery (DO₂) and consumption (VO₂) for the upper body. Overestimates true mixed SvO₂ by ~5–7%.
Twelve questions across all six modules. Click an option to reveal the explanation.
▸ Questions — click to answer
Disclaimer
This resource is created for educational purposes. All content is based on published literature and standard clinical references current to 2025–26. It is not a substitute for institutional guidelines, senior clinical advice, or individual patient assessment.